Bowie Dick Test Pack

Preassembled daily test pack for pre‑vacuum steam sterilizers. Verifies effective air removal and steam penetration across the chamber. Delivers a clear, uniform blue → black color change at the center of the pack after a successful cycle at 134°C / 3.5 min.

  • • ISO 11140‑1 Type 2 compliant indicator sheet
  • • Ready‑to‑use pack—no assembly required
  • • Consistent ink and wicking layer for uniform results
  • • Non‑toxic, heavy‑metal‑free formulation

How to Use

  • Place one test pack in an empty sterilizer chamber on the bottom shelf above the drain (or as per hospital SOP).
  • Run a pre‑vacuum cycle at 134°C for 3.5 minutes (or as per equipment manufacturer guidance).
  • Open the pack and evaluate the indicator sheet. A uniform transition to black indicates a pass; uneven or incomplete change indicates a fail.
  • Record batch, date, sterilizer ID, and operator signature for traceability.

Indicator Result Example

Product Image
Before Sterilization (Blue)
Product Image
After Sterilization (Black)

Ordering Information

Specification Detail Packaging
Preassembled Bowie Dick Test Pack (single‑use), Cycle: 134°C / 3.5 min, Blue → Black BDTP13403 20 Packs
Product Image
Specification Detail
Compliance ISO 11140‑1 Type 2
Format Preassembled pack with indicator sheet
Sterilization Method Steam (pre‑vacuum)
Interpretation Uniform blue → black color change indicates pass
Shelf Life 24 Months
Storage Cool, dry place (15–25°C), away from direct light
Product Image

Enquire Now

We are committed to processing the information in order to contact you and talk about your project.